Home / Recalls / Medical Devices
Remel Inc recalls Oxoid Antimicrobial Susceptibility Testing Disc, MEM10, REF: CT0774B, 10 mcg meropenem, Packed in packs containing 5 ca
Recalled January 22, 2014 · FDA recall Z-0734-2014 · Remel
Hazard
Individual discs in the lot may not be sufficiently impregnated with antibiotic, resulting in smaller than expected zone sizes. · FDA Class II
Units
45/5/50 - discs/cartridges/packs
Sold at
Nationwide Distribution including the states of AL, AZ, CA, CO, FL, HI, IA, IL, KY, LA, MD, MI, MN, MO, MS, NC, NY, OH, SC, TN, TX, VA, WI, and WV
Description
Individual discs in the lot may not be sufficiently impregnated with antibiotic, resulting in smaller than expected zone sizes.
Product
Oxoid Antimicrobial Susceptibility Testing Disc, MEM10, REF: CT0774B, 10 mcg meropenem, Packed in packs containing 5 cartridges per pack, 50 discs per cartridge. Firm name on the label is Oxoid, Ltd., Basingstoke, United Kingdom. Used in the semi-quantitative agar diffusion test method for in vitro susceptibility testing to assist in the determination of a suitable antibiotic agent for treatment purposes.
Manufacturer(s)
Remel Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →