Reshape Medical Inc recalls ReShape Integrated Dual Balloon System, Gen 1, Catalog No. RSM110, Model No. 01-0013-001 The ReShape Integrated Dual
Recalled August 24, 2016 · FDA recall Z-2552-2016 · Reshape Medical
Hazard
Due to a suspected misalignment of the connection between the proximal balloon cap of the implant and the delivery catheter. The possible misalignment is the result of an assembly error during manufacturing and may lead to leakage of saline solution during the balloon inflation procedure. · FDA Class II
Units
56 units
Sold at
US: Distribution to the states of : NY, OH and CA.
Description
Due to a suspected misalignment of the connection between the proximal balloon cap of the implant and the delivery catheter. The possible misalignment is the result of an assembly error during manufacturing and may lead to leakage of saline solution during the balloon inflation procedure.
Product
ReShape Integrated Dual Balloon System, Gen 1, Catalog No. RSM110, Model No. 01-0013-001 The ReShape Integrated Dual Balloon System is indicated for weight reduction when used in conjunction with diet and exercise, in obese patients with a Body Mass Index (BMI) of 30 40 kg/m2 and one or more obesity related comorbid conditions. It is indicated for use in adult patients who have failed weight reduction with diet and exercise alone.
Manufacturer(s)
Reshape Medical Inc