RecallWatchUSA

Home / Recalls / Medical Devices

Respironics California Inc recalls Philips V60 Ventilators, Material No. 850008, 1053613, 1053614, 1053615, 1053616, 1053617, 1076715, 1076716, 1076717, DU

Recalled May 24, 2017 · FDA recall Z-2061-2017 · Respironics California

Hazard
Over time, low-frequency vibrations can cause the pins within the female connectors on the Motor Controller to Data Acquisition Board ribbon cable to become partially displaced causing momentary high resistance that prevents or restricts data travelling through the cable during the high resistance event. · FDA Class I
Units
30,516 units
Sold at
Worldwide Distribution- US (nationwide) and countries of: Argentina, Australia, Austria, Bahrain, Barbados, Belgium, Bolivia, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Hong Kong, Iceland, India, Indonesia, Iran (Islamic Republic Of), Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Korea, Republic Of, Kuwait, Lebanon, Libyan Arab Jamahiriya, Malaysia, Malta, Mexico, Monaco, Morocco, Namibia, Netherlands, Netherlands Antilles, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Province Of China, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Viet Nam.

Description

Over time, low-frequency vibrations can cause the pins within the female connectors on the Motor Controller to Data Acquisition Board ribbon cable to become partially displaced causing momentary high resistance that prevents or restricts data travelling through the cable during the high resistance event.

Product
Philips V60 Ventilators, Material No. 850008, 1053613, 1053614, 1053615, 1053616, 1053617, 1076715, 1076716, 1076717, DU1053617, U1053617. Old Material No. 850008, 989805628251, 989805612101, 989805613391, 989805613661, 989805611761, 989805627411, 989805627431, 989805627441, 989805616411, 989805636631.
Manufacturer(s)
Respironics California Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Respironics California Inc recalls Philips V60 Ventilators, Material No. 850008, 1053613, 1053614, 1053615, 1053616, 1053617, 1076715, 1076716, 1076717, DU | RecallWatchUSA