Respironics California, LLC recalls Respironics V60 Ventilator, Service # 850008, REF: 1053613, 1053614, 1053617, 1053614
Recalled February 19, 2020 · FDA recall Z-1152-2020 · Respironics California
Hazard
Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchscreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia. · FDA Class II
Units
2851
Sold at
US Nationwide. OUS (Foreign): Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Indonesia, Israel, Italy, Malaysia, Mexico, Netherlands, Portugal, Singapore, South Africa, Spain, Sweden, United Arab Emirates, United Kingdom, Viet Nam, Oman, Philippines, Saudi Arabia, Taiwan, Turkey, Russian Federation, Argentina, Ecuador, Poland, Guatemala, Peru, Viet Nam
Description
Ventilator touchscreen may become frozen and fail to respond to touch commands. If a patient requires a change in therapy, there is no warning associated with an unresponsive touchscreen, and therefore the clinician would have no prior indication of a touchscreen failure and would be unable to change ventilator settings. The ventilator will continue to function at the predetermined and accepted settings required to support the patient. A delay in therapy adjustments may result in hypercarbia and/or hypoxemia.
Product
Respironics V60 Ventilator, Service # 850008, REF: 1053613, 1053614, 1053617, 1053614
Manufacturer(s)
Respironics California, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →