RecallWatchUSA

Home / Recalls / Medical Devices

Riverpoint Medical, LLC recalls Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S

Recalled December 20, 2023 · FDA recall Z-0518-2024 · Riverpoint Medical

Hazard
Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use. · FDA Class II
Units
999 devices
Sold at
U.S. Nationwide distribution in the state of CA.

Description

Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.

Product
Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 3911-714-521 The Iconix Knotless anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder.
Manufacturer(s)
Riverpoint Medical, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov