Home / Recalls / Medical Devices
Roche Diagnostics Corporation recalls cobas e 411 Immunoassay Analyzer (Discrete Photometric Analyzer for Clinical use), Catalog Numbers: 04775201001 (rack s
Recalled April 4, 2018 · FDA recall Z-1270-2018 · Roche Diagnostics
Hazard
Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 mm sample tubes without tube adapters. · FDA Class II
Sold at
US Nationwide
Description
Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 mm sample tubes without tube adapters.
Product
cobas e 411 Immunoassay Analyzer (Discrete Photometric Analyzer for Clinical use), Catalog Numbers: 04775201001 (rack system), 04775279001 (disk system), 04775201973 (rack system) & 04775279973 (disk system).
Manufacturer(s)
Roche Diagnostics Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →