Roche Diagnostics Corporation recalls Cobas¿ infinity central lab IT solution Material Number: 07154003001 Calculator/Data Processing Module, For Clinic

Recalled May 29, 2019 · FDA recall Z-1368-2019 · Roche Diagnostics

Hazard
Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the results from the first sample subsequent to the bottle change, (when using the Standby bottle function), does not display a quality control (QC) alarm when one should be present. If the QC is out of range on the stand-by bottle and it is not removed from the system, there is a potential that when the status of the Standby bottle changes from the Standby bottle to the Current bottle status, QC alarms will not be displayed next to the sample result. · FDA Class II
Units
15 units
Sold at
US Distribution to states of: AL, AZ, CA, GA, IA, MO, NJ, NY, SC, TN and TX.

Description

Using the following versions of cobas infinity software (2.0 thorough 2.5), there is a potential where the results from the first sample subsequent to the bottle change, (when using the Standby bottle function), does not display a quality control (QC) alarm when one should be present. If the QC is out of range on the stand-by bottle and it is not removed from the system, there is a potential that when the status of the Standby bottle changes from the Standby bottle to the Current bottle status, QC alarms will not be displayed next to the sample result.

Product
Cobas¿ infinity central lab IT solution Material Number: 07154003001 Calculator/Data Processing Module, For Clinical Use
Manufacturer(s)
Roche Diagnostics Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov