RecallWatchUSA

Home / Recalls / Medical Devices

Roche Diagnostics Corporation recalls Lactate Gen.2, Catalog Nos. 03183700190, 05171881190, 11822837190, ***added as of 5/10/19*** 03183700190 Product

Recalled April 24, 2019 · FDA recall Z-1146-2019 · Roche Diagnostics

Hazard
New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS INTEGRA 400 plus analyzers are also impacted for BILT3 and LACT2 assays*** · FDA Class II
Units
5292 analytical units
Sold at
US Nationwide distribution.

Description

New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS INTEGRA 400 plus analyzers are also impacted for BILT3 and LACT2 assays***

Product
Lactate Gen.2, Catalog Nos. 03183700190, 05171881190, 11822837190, ***added as of 5/10/19*** 03183700190 Product Usage: An In vitro diagnostic reagent system intended for use on cobas Integra for the quantitative determination of the lactate concentration in plasma and cerebrospinal fluid.
Manufacturer(s)
Roche Diagnostics Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov