Roche Diagnostics Corporation recalls Roche Acetaminophen assay 03255379160-HAcetaminophen (P) 20767174160- cobas c Integra AAcetaminophen150 test

Recalled May 2, 2018 · FDA recall Z-1519-2018 · Roche Diagnostics

Hazard
The previous version of method sheets for the Acetaminophen assay on the COBAS INTEGRA 400 plus/800/800 CTS analyzers and the MODULAR ANALYTICS P module are missing the following statement in the Limitations-interference section: "In very rare cases, gammopathy, in particular type IgM (Waldenstr¿m s macroglobulinemia), may cause unreliable results." · FDA Class II
Units
35,111
Sold at
Nationwide Distribution

Description

The previous version of method sheets for the Acetaminophen assay on the COBAS INTEGRA 400 plus/800/800 CTS analyzers and the MODULAR ANALYTICS P module are missing the following statement in the Limitations-interference section: "In very rare cases, gammopathy, in particular type IgM (Waldenstr¿m s macroglobulinemia), may cause unreliable results."

Product
Roche Acetaminophen assay 03255379160-HAcetaminophen (P) 20767174160- cobas c Integra AAcetaminophen150 test
Manufacturer(s)
Roche Diagnostics Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Roche Diagnostics Corporation recalls Roche Acetaminophen assay 03255379160-HAcetaminophen (P) 20767174160- cobas c Integra AAcetaminophen150 test | RecallWatchUSA