Roche Diagnostics Operations, Inc. recalls ACCU-CHEK Aviva Combo Blood Glucose Meter (as a part of the ACCU-CHEK Aviva Combo System)\ The ACCU-CHEK Combo System

Recalled December 4, 2013 · FDA recall Z-0385-2014 · Roche Diagnostics Operations

Hazard
There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pump option is chosen on the ACCU-CHEK Aviva Combo meter, there is a possibility of receiving an incorrect bolus advice recommendation that may cause a temporary under delivery of insulin. · FDA Class II
Units
6,595 units
Sold at
Nationwide Distribution.

Description

There is a software synchronization issue with the ACCU-CHEK Combo system. In rare cases, when the Manual Pump option is chosen on the ACCU-CHEK Aviva Combo meter, there is a possibility of receiving an incorrect bolus advice recommendation that may cause a temporary under delivery of insulin.

Product
ACCU-CHEK Aviva Combo Blood Glucose Meter (as a part of the ACCU-CHEK Aviva Combo System)\ The ACCU-CHEK Combo System is indicated for the treatment of insulin-requiring diabetes and for the quantitative measurement of glucose in fresh capillary whole blood from the finger.
Manufacturer(s)
Roche Diagnostics Operations, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov