Roche Diagnostics Operations, Inc. recalls ACCU-CHEK Tender I 13/80 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body
Recalled January 7, 2015 · FDA recall Z-0884-2015 · Roche Diagnostics Operations
Hazard
Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify · FDA Class II
Units
Total of 13,041 boxes
Sold at
US Nationwide Distribution
Description
Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify
Product
ACCU-CHEK Tender I 13/80 10 pieces Product Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a proprietary click-lock connector for the infusion of drugs subcutaneously. The infusion set can be detached from the indwelling catheter, and the catheter capped to allow freedom from the infusion set and pump for showers, athletics or other activities.
Manufacturer(s)
Roche Diagnostics Operations, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →