Roche Diagnostics Operations, Inc. recalls Cholesterol CHOD-PAP. In vitro diagnostic reagent system intended for use on the Hitachi systems for The quantitative
Recalled July 8, 2015 · FDA recall Z-1920-2015 · Roche Diagnostics Operations
Hazard
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine. · FDA Class II
Units
6,255 analytical units distributed
Sold at
US (nationwide) including PR.
Description
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
Product
Cholesterol CHOD-PAP. In vitro diagnostic reagent system intended for use on the Hitachi systems for The quantitative determination of total cholesterol in serum and plasma.
Manufacturer(s)
Roche Diagnostics Operations, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →