RecallWatchUSA

Home / Recalls / Medical Devices

Roche Diagnostics Operations, Inc. recalls cobas infinity central lab, Material Number 07154003001

Recalled January 18, 2023 · FDA recall Z-0952-2023 · Roche Diagnostics Operations

Hazard
A complaint investigation revealed that an incorrect behavior relating to the rejection of orders capability of the Host Connectivity Agent (HCA) could occur where the order received from the Laboratory Information System (LIS) is created with an Internal ID rather than the External ID sent from the LIS. The erroneous event occurs under rare circumstances where the date in the sample's barcode does not match the order date sent from the LIS, and can lead to the order being mismatched to an another patient's sample ID rather than the subject patient's sample ID. The software bug impacts multiple cobas infinity central lab software (versions 2.5.x , 3.01.x, 3.02.x, and 3.03.x) and occurs when the afflicted software has been configured with specific pre-conditions. · FDA Class II
Units
224 US; 2,797 worldwide
Sold at
Domestic distribution nationwide. Foreign distribution pending.

Description

A complaint investigation revealed that an incorrect behavior relating to the rejection of orders capability of the Host Connectivity Agent (HCA) could occur where the order received from the Laboratory Information System (LIS) is created with an Internal ID rather than the External ID sent from the LIS. The erroneous event occurs under rare circumstances where the date in the sample's barcode does not match the order date sent from the LIS, and can lead to the order being mismatched to an another patient's sample ID rather than the subject patient's sample ID. The software bug impacts multiple cobas infinity central lab software (versions 2.5.x , 3.01.x, 3.02.x, and 3.03.x) and occurs when the afflicted software has been configured with specific pre-conditions.

Product
cobas infinity central lab, Material Number 07154003001
Manufacturer(s)
Roche Diagnostics Operations, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Roche Diagnostics Operations, Inc. recalls cobas infinity central lab, Material Number 07154003001 | RecallWatchUSA