Roche Diagnostics Operations, Inc. recalls RSD 800 Sorting System calculator/data processing module, for clinical use Product Usage: The cobas 8000 Modular An

Recalled November 2, 2016 · FDA recall Z-0105-2017 · Roche Diagnostics Operations

Hazard
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process. · FDA Class II
Units
9014 in total
Sold at
US Nationwide Distribution

Description

Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

Product
RSD 800 Sorting System calculator/data processing module, for clinical use Product Usage: The cobas 8000 Modular Analyze Series is a fully automated system for Clinical chemistry analysis, intended for the in vitro qualitative and quantitative determination of analytes in body fluids. It is optimized for high throughput workloads in the professional environment using a combination of ion selective electrodes (ISE) and a photometric analysis unit.
Manufacturer(s)
Roche Diagnostics Operations, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov