Roche Diagnostics Operations, Inc. recalls RSD 800A Aliquoting System calculator/data processing module, for clinical use Product Usage: The Roche/Hitachi MOD

Recalled November 2, 2016 · FDA recall Z-0104-2017 · Roche Diagnostics Operations

Hazard
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process. · FDA Class II
Units
9014 in total
Sold at
US Nationwide Distribution

Description

Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.

Product
RSD 800A Aliquoting System calculator/data processing module, for clinical use Product Usage: The Roche/Hitachi MODULAR Analytics System is a fully automated, random-access, software-controlled system for immunoassay and photometric analysis intended for qualitative and quantitative in vitro determinations using a wide variety of tests.
Manufacturer(s)
Roche Diagnostics Operations, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov