Roche Molecular Systems, Inc. recalls cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondar
Recalled May 29, 2013 · FDA recall Z-1346-2013 · Roche Molecular Systems
Hazard
In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported · FDA Class II
Units
425 instruments
Sold at
Nationwide Distribution.
Description
In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported
Product
cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.
Manufacturer(s)
Roche Molecular Systems, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →