RecallWatchUSA

Home / Recalls / Medical Devices

Roche Molecular Systems, Inc. recalls cobas¿ Liat¿ Analyzer

Recalled July 27, 2016 · FDA recall Z-2216-2016 · Roche Molecular Systems

Hazard
Device was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid results. · FDA Class II
Units
2
Sold at
Distributed in NC and WA.

Description

Device was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid results.

Product
cobas¿ Liat¿ Analyzer
Manufacturer(s)
Roche Molecular Systems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov