Roche Molecular Systems, Inc. recalls MagNA Pure 96 Instrument, clinical sample concentrator, Material number 06541089001 - Product Usage: This protocol is i

Recalled July 22, 2020 · FDA recall Z-2551-2020 · Roche Molecular Systems

Hazard
When using Sample Transfer protocol version 3.0, the drop catcher is not activated on the MagNA Pure 96 instrument, which introduces a potential risk for cross-contamination leading to erroneous results. · FDA Class II
Units
917
Sold at
Worldwide distribution - US Nationwide distribution.

Description

When using Sample Transfer protocol version 3.0, the drop catcher is not activated on the MagNA Pure 96 instrument, which introduces a potential risk for cross-contamination leading to erroneous results.

Product
MagNA Pure 96 Instrument, clinical sample concentrator, Material number 06541089001 - Product Usage: This protocol is intended for pipetting samples into a MagNA Pure 96 Processing Cartridge (purification rack position 1) before a purification run.
Manufacturer(s)
Roche Molecular Systems, Inc.