RecallWatchUSA

Home / Recalls / Medical Devices

Roche Molecular Systems, Inc. recalls uPath software 1.1 and uPath software 1.1 IVD-INSW, Material numbers 08991871001 and 09077332001

Recalled January 27, 2021 · FDA recall Z-0877-2021 · Roche Molecular Systems

Hazard
When a user creates a measurement tool annotation in the uPath Enterprise software version 1.1, the measurement value is incorrectly calculated when the measurement is viewed in "Split View" viewing mode and the slides within the case are scanned at different magnifications. · FDA Class II
Units
27
Sold at
Domestic distribution to Maryland and Michigan. International distribution to France, Greece, Italy, Poland, Turkey, Brazil, India, and Singapore.

Description

When a user creates a measurement tool annotation in the uPath Enterprise software version 1.1, the measurement value is incorrectly calculated when the measurement is viewed in "Split View" viewing mode and the slides within the case are scanned at different magnifications.

Product
uPath software 1.1 and uPath software 1.1 IVD-INSW, Material numbers 08991871001 and 09077332001
Manufacturer(s)
Roche Molecular Systems, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Roche Molecular Systems, Inc. recalls uPath software 1.1 and uPath software 1.1 IVD-INSW, Material numbers 08991871001 and 09077332001 | RecallWatchUSA