Sarnova HC, Llc recalls Curaplex Field Cric Kit with ET Tube, Item Number 36611. The kit provides healthcare professionals with an assortment o

Recalled January 24, 2024 · FDA recall Z-0720-2024 · Sarnova HC

Hazard
It was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube. · FDA Class II
Units
1063 kits
Sold at
US Nationwide

Description

It was determined that the Endotracheal Tube Holder included in certain kits is not compatible with a cricothyrotomy procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.

Product
Curaplex Field Cric Kit with ET Tube, Item Number 36611. The kit provides healthcare professionals with an assortment of supplies to perform a cricothyrotomy.
Manufacturer(s)
Sarnova HC, Llc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov