Home / Recalls / Medical Devices
Sarnova HC, Llc recalls Curaplex Nasal Intubation Kit with 6mm Endotrol Tube-To perform a nasal intubation to establish an airway in an emergenc
Recalled March 27, 2024 · FDA recall Z-1307-2024 · Sarnova HC
Hazard
Endotracheal Tube Holder included in certain kits is not compatible with a nasal intubation procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube. · FDA Class II
Units
540 kits
Sold at
US Nationwide distribution.
Description
Endotracheal Tube Holder included in certain kits is not compatible with a nasal intubation procedure, leaving practitioners without the necessary means to secure the inserted endotracheal tube.
Product
Curaplex Nasal Intubation Kit with 6mm Endotrol Tube-To perform a nasal intubation to establish an airway in an emergency. Item Number: 023060S
Manufacturer(s)
Sarnova HC, Llc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →