RecallWatchUSA

Home / Recalls / Medical Devices

Scientia Vascular, Inc. recalls Plato 17 Microcatheter: Straight Tip, PL17-160-000 ; 45 degree Tip, PL17-160-045; 90 degree Tip, PL17-160-090

Recalled March 22, 2023 · FDA recall Z-1249-2023 · Scientia Vascular

Hazard
Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response. · FDA Class II
Units
353
Sold at
US Nationwide distribution in the states of IL, AL, FL, NC, WI, GA, SC, IN, KS, CA, OH, NY, NH, CO, MA, NV, UT, TX.

Description

Microcatheters for embolic coil and diagnostic agent introduction to peripheral/neuro vasculature, due to manufacturing non-conformance, and due to flush, guidewire insertion and manipulation prior to placement, may have loose particles within catheter lumens that may be introduced to patients, which could cause embolic event, small vessel occlusion, and subacute to chronic inflammatory response.

Product
Plato 17 Microcatheter: Straight Tip, PL17-160-000 ; 45 degree Tip, PL17-160-045; 90 degree Tip, PL17-160-090
Manufacturer(s)
Scientia Vascular, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov