Home / Recalls / Medical Devices
Securitas Healthcare LLC recalls Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3
Recalled January 14, 2026 · FDA recall Z-1031-2026 · Securitas Healthcare
Hazard
Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing. · FDA Class II
Units
5627 units
Sold at
Worldwide distribution -US Nationwide and the country of Canada.
Description
Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.
Product
Arial 900 MHz Call Station, Model Number CSK200-900, with Firmware version: 1.6.3
Manufacturer(s)
Securitas Healthcare LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →