Seiler Instr. & Mfg. Co. Inc. recalls Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.

Recalled April 11, 2018 · FDA recall Z-1294-2018 · Seiler Instr. & Mfg

Hazard
The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post. · FDA Class II
Units
43 units, but 1 affected
Sold at
Worldwide Distribution - US Distribution to the states of : CA, CO, FL, IL, MN, NV, OK, TX., and to the countries of : Canada, China, Ecuador, England, Greece, Guatemala, Jordan, Lebanon, Mexico, Romania, Russia, South Korea, Taiwan, Japan and United Arab Emirates.

Description

The microscope may tip over if attachments are fully extended in one line and there is an attempt to move the device not by the handles mounted on the post.

Product
Alpha Air 6 ENT Microscope Microscope examination of the human ear, nose, and throat.
Manufacturer(s)
Seiler Instr. & Mfg. Co. Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov