SenTec AG recalls SenTec Membrane Changer Insert [5 pcs - Product Usage: is indicated for continuous, non-invasive monitoring of carbon di

Recalled August 14, 2019 · FDA recall Z-2156-2019 · SenTec AG

Hazard
There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure. · FDA Class II
Units
205 pcs
Sold at
Worldwide distribution - US Nationwide in the state of Missouri and countries of Argentina, Australia, Austria, Belgium, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, Ecuador, Europe, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Indonesia, Israel, Italy, Kuwait, Malaysia, Finland, Portugal, Romania, Russia, Saudi Arabia, Singapore, South Korea, Spain, Sri Lanka, Switzerland, Taiwan, Thailand, United Arab Emirates and United Kingdom.

Description

There is an increased chance (approx. 5%) of no electrolyte or a reduced amount of electrolyte being applied to the sensor during the membrane change procedure.

Product
SenTec Membrane Changer Insert [5 pcs - Product Usage: is indicated for continuous, non-invasive monitoring of carbon dioxide tension and oxygen tension as well as oxygen saturation and pulse rate in adult and pediatric patients. In neonatal patients the SDMS is indicated for carbon dioxide and oxygen tension monitoring only. Oxygen tension monitoring is contraindicated for patients under gas anesthesia.]
Manufacturer(s)
SenTec AG
Made in
Switzerland

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov