SentreHeart Inc recalls FindrWIRZ Guidewire System Catalog# 40-05 Cardiology: The system is intended for use in the cardiovascular system fo
Recalled November 16, 2016 · FDA recall Z-0328-2017 · SentreHeart
Hazard
A complaint was received on September 27th, 2016 at a medical center in Germany for delamination and flaking of the PTFE lubricious coating on three FindrWIRZ Guidewire System .025 guidewires removed from their packaging hoops. · FDA Class I
Units
132 worldwide distribution (98) of those units in US
Sold at
Worldwide Distribution - US including MO, CA, OR, GA. NH, NY, MI, TX, MD, PA, IL, KS, NE, CT, AZ and Internationally to United Kingdom and Germany
Description
A complaint was received on September 27th, 2016 at a medical center in Germany for delamination and flaking of the PTFE lubricious coating on three FindrWIRZ Guidewire System .025 guidewires removed from their packaging hoops.
Product
FindrWIRZ Guidewire System Catalog# 40-05 Cardiology: The system is intended for use in the cardiovascular system for introduction and positioning of over the-wire catheters and therapeutic devices during interventional procedures.
Manufacturer(s)
SentreHeart Inc