Home / Recalls / Medical Devices
Separation Technology, Inc. recalls SLIDERPREP PLUS SLIDE PREPARATION CENTRIFUGE. Produce a monolayer of cells onto a glass slide from any fluid suspensi
Recalled May 29, 2013 · FDA recall Z-1367-2013 · Separation Technology
Hazard
Separation Technology, Inc. recalled SlidePrep Plus Cytology Centrifuges manufactured prior to January 1, 2013 for having the potential to start up unexpectedly when encountering an electrostatic discharge. · FDA Class II
Units
102
Sold at
Worldwide Distribution-USA (nationwide) including the states of PA, NC, IL, KS, and FL, and the countries Canada and Turkey.
Description
Separation Technology, Inc. recalled SlidePrep Plus Cytology Centrifuges manufactured prior to January 1, 2013 for having the potential to start up unexpectedly when encountering an electrostatic discharge.
Product
SLIDERPREP PLUS SLIDE PREPARATION CENTRIFUGE. Produce a monolayer of cells onto a glass slide from any fluid suspension.
Manufacturer(s)
Separation Technology, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →