Shanghai United Imaging Healthcare Co., Ltd. recalls uEXPLORER PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthca
Recalled August 19, 2020 · FDA recall Z-2799-2020 · Shanghai United Imaging Healthcare
Hazard
1) 1) Potential intermittent issue may cause image data problems, may lead to artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions. · FDA Class II
Units
1 unit distributed to Texas
Sold at
US Nationwide distribution including in the state of Texas.
Description
1) 1) Potential intermittent issue may cause image data problems, may lead to artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions.
Product
uEXPLORER PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
Manufacturer(s)
Shanghai United Imaging Healthcare Co., Ltd.
Made in
China