RecallWatchUSA

Home / Recalls / Medical Devices

Shimadzu Corp. Analytical & Measuring Instruments Div. recalls Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL

Recalled December 4, 2024 · FDA recall Z-0553-2025 · Shimadzu Corp. Analytical & Measuring Instruments Div

Hazard
There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing. · FDA Class II
Units
179
Sold at
Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.

Description

There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.

Product
Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL
Manufacturer(s)
Shimadzu Corp. Analytical & Measuring Instruments Div.
Made in
Japan

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Shimadzu Corp. Analytical & Measuring Instruments Div. recalls Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL | RecallWatchUSA