Home / Recalls / Medical Devices
Shimadzu Medical Systems recalls 0.4/0.7JG326D-265AX/AT Product Usage: The device is a mobile X-ray system in order to take general radiography of
Recalled November 14, 2012 · FDA recall Z-0265-2013 · Shimadzu Medical Systems
Hazard
Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006. · FDA Class II
Units
3 units
Sold at
US Nationwide Distribution
Description
Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.
Product
0.4/0.7JG326D-265AX/AT Product Usage: The device is a mobile X-ray system in order to take general radiography of patients in the hospital who cannot move and/or of outpatients in emergency.
Manufacturer(s)
Shimadzu Medical Systems
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →