Shimadzu Medical Systems recalls 0.7U163CS-36 In combination with below systems MUX-100H, MUX-100D, MobileArt Evolution Brand Name: MUX-100H. Cl

Recalled November 14, 2012 · FDA recall Z-0262-2013 · Shimadzu Medical Systems

Hazard
Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006. · FDA Class II
Units
288 units
Sold at
US Nationwide Distribution

Description

Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.

Product
0.7U163CS-36 In combination with below systems MUX-100H, MUX-100D, MobileArt Evolution Brand Name: MUX-100H. Classification Name: Mobile x-ray system. Device Model#MUX-100H. Product Usage: The device is a mobile X-ray system in order to take general radiography of patients in the hospital who cannot move and/or of outpatients in emergency.
Manufacturer(s)
Shimadzu Medical Systems

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Shimadzu Medical Systems recalls 0.7U163CS-36 In combination with below systems MUX-100H, MUX-100D, MobileArt Evolution Brand Name: MUX-100H. Cl | RecallWatchUSA