Home / Recalls / Medical Devices
Shimadzu Medical Systems recalls Bransist Safire, Digital Angiography System containing a MS-200 Ceiling Suspended C-arm Support - Product Usage: designe
Recalled December 16, 2020 · FDA recall Z-0554-2021 · Shimadzu Medical Systems
Hazard
The firm has identified a problem with the control software for the celling arm of the X-Ray System . This software issue could cause the ceiling arm of the device to move after the operator has released the movement lever. This could result in possible collision or harm to the patient or system operator. · FDA Class II
Units
7 units
Sold at
US Nationwide distribution including in the states of KY, CA, MI, OH, CT, MI, NY, SC, NC,IN, IL, TN, LA, CT, MT.
Description
The firm has identified a problem with the control software for the celling arm of the X-Ray System . This software issue could cause the ceiling arm of the device to move after the operator has released the movement lever. This could result in possible collision or harm to the patient or system operator.
Product
Bransist Safire, Digital Angiography System containing a MS-200 Ceiling Suspended C-arm Support - Product Usage: designed to provide fluoroscopic images of the patient during diagnostic surgical and interventional angiography and cardiology procedures.
Manufacturer(s)
Shimadzu Medical Systems
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →