RecallWatchUSA

Home / Recalls / Medical Devices

Shimadzu Medical Systems recalls Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40 / D150GC-

Recalled June 10, 2026 · FDA recall Z-2294-2026 · Shimadzu Medical Systems

Hazard
Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off. · FDA Class II
Units
3487 units
Sold at
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Canada, The Bahamas, China, Denmark, Haiti.

Description

Electromagnetic contactors within high voltage generator may weld together, resulting in current being supplied to power supply even if power is off.

Product
Digital Angiography System Trinias:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST safire:UD150G-40 / D150GC-40; Digital Angiography System BRANSIST alexa:UD150G-40 / D150GC-40; Digital Radiography System RADspeed Pro: UD150B-40 / UD150V-40 / UD150L-40 / UD150L-40E/D150BC-40 / D150VC-40 / D150LC-40 / D150LC-40E/D150BC-41 / D150VC-41 / D150LC-41; Digital Radiography System RADspeed safire: UD150B-40 / D150BC-40; Radiography System RADspeed fit: N/A (Integrated in the system); X-RAY R/F SYSTEM SONIALVISION safire17:UD150B-40 / D150BC-40; X-RAY R/F SYSTEM SONIALVISION safire Plus: UD150B-40 / D150BC-4;0 X-RAY R/F SYSTEM SONIALVISION safire: UD150B-40 / D150BC-40; X-RAY R/F SYSTEM SONIALVISION VERSA: UD150B-40 / UD150V-40 / UD150L-40/D150BC-40 / D150VC-40 / D150LC-40; X-RAY R/F SYSTEM FLUOROspeed 300:UD150B-40 / UD150V-40 / UD150L-40/D150BC-40 / D150VC-40 / D150LC-40; X-RAY R/F SYSTEM SONIALVISION G4:D150BC-40S; X-RAY R/F SYSTEM FLUOROspeed X1:D150BC-40S
Manufacturer(s)
Shimadzu Medical Systems

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov