RecallWatchUSA

Home / Recalls / Medical Devices

Shimadzu Medical Systems recalls R-20J , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-Ray tubes: 06/1.2P364DK-85, 0.6/1

Recalled November 14, 2012 · FDA recall Z-0261-2013 · Shimadzu Medical Systems

Hazard
Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006. · FDA Class II
Units
308 units
Sold at
US Nationwide Distribution

Description

Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.

Product
R-20J , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-Ray tubes: 06/1.2P364DK-85, 0.6/1.2P324DK-125, 06/1.2P324DK-85, 0.6/1.2P38DE-85. Product Usage: The device is the beam-limiting device for X-ray Radiography.
Manufacturer(s)
Shimadzu Medical Systems

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Shimadzu Medical Systems recalls R-20J , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-Ray tubes: 06/1.2P364DK-85, 0.6/1 | RecallWatchUSA