Shimadzu Medical Systems Usa Com recalls SHIMADZU Ceiling Type X-ray Tube Support CH-200/CH-200M Catalog Number: CH-200/CH-200M Holds an x-ray tube unit and col

Recalled July 5, 2017 · FDA recall Z-2653-2017 · Shimadzu Medical Systems Usa Com

Hazard
Cracks may occur over time on the tube holding shaft near the base of the tube mounting flange. · FDA Class II
Units
1673 units total
Sold at
Worldwide Distribution - Nationwide distribution to AK, AR, AZ, CA, CT, FL, IL, IN, KS, LA, MI, MO, NE, NJ, NM, NV, NY,OH, OK, PA, RI, SC, TN, TX, WA, WI, and Hawaii Foreign: Canada and Japan

Description

Cracks may occur over time on the tube holding shaft near the base of the tube mounting flange.

Product
SHIMADZU Ceiling Type X-ray Tube Support CH-200/CH-200M Catalog Number: CH-200/CH-200M Holds an x-ray tube unit and collimator combined with an z-ray high voltage generator, radiographic stand, and table. It is designed to perform radiography of patients either in a standing or supine position
Manufacturer(s)
Shimadzu Medical Systems Usa Com