Siemens Healthcare Diagnostics recalls ADVIA Centaur XP Immunoassay System (including refurbished units), This system is intended for professional use in a

Recalled October 8, 2014 · FDA recall Z-2716-2014 · Siemens Healthcare Diagnostics

Hazard
The firm has identified an issue with patient demographic information sent to the LIS from the ADVIA Centaur/ADVIA Centaur XP Immunoassay systems. Siemens has confirmed that under extremely rare circumstances patient demographic data from the previous order received from the LIS is merged with the next order. This issue can occur when the LIS data buffer on the ADVIA Centaur system becomes full · FDA Class II
Units
Domestic: 2771 units; Foreign: 7744 units
Sold at
Worldwide Distribution including the United States., and the countries of Argentina, Austria, Australia, Bahrain, Bangladesh, Belgium, Belgrade, Brazil, Botswana, Canada, Chile, China, Colombia, Croatia, Czech Republic, Cyprus, Denmark, Egypt, Estonia, Finland, France, Germany, Georgia, Great Britain, Hungary, Indonesia, India, Ireland, Israel, Italy, Japan, Jordan, Korea, Kazakhstan, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Malta, Martinique, Mexico, Myanmar, New Caledonia, Netherlands, Norway, New Zealand, Pakistan, French Polynesia, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, Slovakia, South Africa, Spain, Switzerland, Thailand, Turkey, United Arab Emirates, Vatican, Venezuela, Vietnam, Yemen and Zimbabwe.

Description

The firm has identified an issue with patient demographic information sent to the LIS from the ADVIA Centaur/ADVIA Centaur XP Immunoassay systems. Siemens has confirmed that under extremely rare circumstances patient demographic data from the previous order received from the LIS is merged with the next order. This issue can occur when the LIS data buffer on the ADVIA Centaur system becomes full

Product
ADVIA Centaur XP Immunoassay System (including refurbished units), This system is intended for professional use in a laboratory environment only. Tests performed using this system are intended for in vitro diagnostic use. As with all diagnostic tests, a definitive clinical diagnosis should not be based on the results of a single test, but should only be made by the physician after all clinical and laboratory findings have been evaluated.
Manufacturer(s)
Siemens Healthcare Diagnostics

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Siemens Healthcare Diagnostics recalls ADVIA Centaur XP Immunoassay System (including refurbished units), This system is intended for professional use in a | RecallWatchUSA