Siemens Healthcare Diagnostics recalls IMMULITE 2000/2000 XPi Systems Androstenedione (Test Code AND), REF L2KAO2, SMN 10381188, IVD, 200 test kit size --- Si

Recalled May 7, 2014 · FDA recall Z-1532-2014 · Siemens Healthcare Diagnostics

Hazard
Siemens Healthcare Diagnostics confirmed over-recovery of samples with concentrations greater than 5.5 ng/mL with the IMMULITE/IMMULITE 1000 and IMMULITE 2000/IMMULITE 2000 XPi Androstenedione assays (LKAO1, L2KAO2). Siemens has determined that these lots are linear up to a concentration of 5.5 ng/mL, instead of 10 ng/mL as stated in the Instructions for Use (IFU). Quality Control materials at concentrations between 5.5 and 10 ng/mL will detect this issue. · FDA Class II
Units
7741
Sold at
Worldwide distribution: US (nationwide) and countries of: Argentina, Australia, Austria, Belgium, Brazil, Canada, China (People's Republic of), Czech.Rebuplik, Denmark, Estonia, France, Germany, Great Britain, Italy, New Zealand, Portugal, Spain, Switzerland, Turkey and Vatican City.

Description

Siemens Healthcare Diagnostics confirmed over-recovery of samples with concentrations greater than 5.5 ng/mL with the IMMULITE/IMMULITE 1000 and IMMULITE 2000/IMMULITE 2000 XPi Androstenedione assays (LKAO1, L2KAO2). Siemens has determined that these lots are linear up to a concentration of 5.5 ng/mL, instead of 10 ng/mL as stated in the Instructions for Use (IFU). Quality Control materials at concentrations between 5.5 and 10 ng/mL will detect this issue.

Product
IMMULITE 2000/2000 XPi Systems Androstenedione (Test Code AND), REF L2KAO2, SMN 10381188, IVD, 200 test kit size --- Siemens Healthcare Diagnostics Products Ltd. Llanberis, Gwynedd, LL55 4EL, UK For in vitro diagnostic use with the IMMULITE/IMMULITE 1000 and 2000 Systems Analyzers for the quantitative measurement of 4-androstenedione in human serum.
Manufacturer(s)
Siemens Healthcare Diagnostics