Home / Recalls / Medical Devices
Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur¿ Cyclosporine
Recalled May 16, 2018 · FDA recall Z-1687-2018 · Siemens Healthcare Diagnostics
Hazard
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere · FDA Class II
Units
N/A
Sold at
Nationwide and Canada, Mexico, and Thailand
Description
The concentrations for biotin listed in the non-interfering substances section of the instructions for use for the Dimension TSHL, Dimension LOCI FT4L, Dimension Vista TSH, ADVIA Centaur Cyclosporine, and ADVIA Centaur DHEAS, incorrectly state the level at which biotin does not interfere
Product
ADVIA Centaur¿ Cyclosporine
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →