Siemens Healthcare Diagnostics, Inc. recalls ADVIA Centaur¿ Folate (FOL)

Recalled May 16, 2018 · FDA recall Z-1689-2018 · Siemens Healthcare Diagnostics

Hazard
Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin. · FDA Class II
Units
N/A
Sold at
Nationwide and Canada, Mexico, and Thailand

Description

Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista MMB, Dimension Sirolimus, ADVIA Centaur FOL, ADVIA Centaur aHBcM and the following IMMULITE assays, 3gAllergy Specific IgE, anti-HBc, BR-MA (CA15-3), CEA, Folic Acid, Gastrin, OM-MA (CA125), Thyroglobulin, and Vitamin B12, and these assays are susceptible to interference from biotin.

Product
ADVIA Centaur¿ Folate (FOL)
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.