Siemens Healthcare Diagnostics, Inc. recalls Advia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the followin

Recalled November 23, 2016 · FDA recall Z-0589-2017 · Siemens Healthcare Diagnostics

Hazard
Siemens has confirmed that the DBIL_2 Assigned Calibrator Values (ACV) currently assigned to the affected lots may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the reference method for Direct Bilirubin measurement. · FDA Class II
Units
10,614 units
Sold at
Worldwide Distribution - US (Nationwide) Internationally to Canada, Mexico, Algeria, Australia, Austria, Bangladesh, Belgium, Brazil, Chile, Cyprus, Denmark, Egypt, Ecuador, Finland, France, Germany, Great Britain, Hong Kong, Hungary, Italy, Japan, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Pakistan, Paraguay, Peru, Poland, Portugal, Russian Fed. Slovakia, Slovenia, Spain, Sweden, Switzerland, Thailand, Turkey, Vatican City and Vietnam

Description

Siemens has confirmed that the DBIL_2 Assigned Calibrator Values (ACV) currently assigned to the affected lots may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the reference method for Direct Bilirubin measurement.

Product
Advia Chemistry Calibrator For in vitro diagnostic use in the calibration of ADVIA¿ Chemistry systems for the following methods: Albumin (ALB, ALB_c), Direct Bilirubin (DBIL_2), Total Bilirubin (TBIL_2), Calcium (CA, CA_c, CA_2, CA_2c), Cholesterol (CHOL), Creatinine (CREA, CREA_2, CRE_2c, ECRE_2), Glucose Hexakinase (GLUH), Glucose Oxidase (GLUO), Iron (IRON_2), Lithium (LITH), Magnesium (MG, MG_c), Phosphorus - inorganic (IP, IP_c), Total Protein (TP, TP_c), Triglycerides (TRIG), Urea Nitrogen (UN, UN_c), Uric Acid (UA, UA_c)