Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AS
Recalled April 12, 2017 · FDA recall Z-1712-2017 · Siemens Healthcare Diagnostics
Hazard
Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferase concentrated (AST_c) assays following calibration with Advia Chemistry Enzyme 2 Calibrator lots 6JD079 and 6MD042. · FDA Class III
Units
314 calibrators
Sold at
Worldwide Distribution - United States Nationwide: Denmark Sweden Norway Great Britain Belgium Netherlands Spain Israel Switzerland France Portugal Greece Germany Slovenia Czech Republic Hong Kong Colombia Korea Australia
Description
Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for Aspartate Aminotransferase (AST) and Aspartate Aminotransferase concentrated (AST_c) assays following calibration with Advia Chemistry Enzyme 2 Calibrator lots 6JD079 and 6MD042.
Product
ADVIA Chemistry Enzyme 2 Calibrator, multi-analyte mixture Product Usage: Assays included in Enzyme 2 Calibrator: AST, AST_c ALT, ALT_c, ALTP5P, ALTP_c, ASTP5P and ASTP_c
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →