Siemens Healthcare Diagnostics, Inc. recalls ADVIA¿ Chemistry (Syva Onboard) Theophylline_2 (THEO_2) Reagent is for in vitro diagnostic use in the quantitative analy

Recalled August 3, 2016 · FDA recall Z-2299-2016 · Siemens Healthcare Diagnostics

Hazard
Siemens Healthcare Diagnostics has confirmed the Syva¿ EMIT¿ 2000 Theophylline reagent lot G3, when run on a Beckman AU Clinical Chemistry System, and the ADVIA¿ Chemistry Theophylline_2 (THEO_2) reagent lot 334018, may exhibit increased imprecision for commercially available Bio-Rad Liquichek TDM Quality Control Levels and patient samples · FDA Class II
Units
658 units
Sold at
Distributed to: CA, CT, DE, FL, GA, IA, IN, KS, KY, LA, MA, MD, MI, MO, NC, NJ, NM, NY, OH, OK, PA, TN, VA, WA

Description

Siemens Healthcare Diagnostics has confirmed the Syva¿ EMIT¿ 2000 Theophylline reagent lot G3, when run on a Beckman AU Clinical Chemistry System, and the ADVIA¿ Chemistry Theophylline_2 (THEO_2) reagent lot 334018, may exhibit increased imprecision for commercially available Bio-Rad Liquichek TDM Quality Control Levels and patient samples

Product
ADVIA¿ Chemistry (Syva Onboard) Theophylline_2 (THEO_2) Reagent is for in vitro diagnostic use in the quantitative analysis of theophylline in human serum or plasma on the ADVIA¿ Chemistry Systems. Copies of IFUs included in submission.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov