RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Healthcare Diagnostics, Inc. recalls Advia Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11319151

Recalled November 3, 2021 · FDA recall Z-0181-2022 · Siemens Healthcare Diagnostics

Hazard
The firm has confirmed the potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely depressed Enzymatic Hemoglobin A1c results may affect consideration of intervention. · FDA Class II
Units
246 (US); 689 (OUS)
Sold at
Worldwide distribution - US Nationwide distribution.

Description

The firm has confirmed the potential for ADVIA Chemistry Urinary/Cerebrospinal Fluid Protein reagent carryover impacting Enzymatic Hemoglobin A1c (A1c_E and A1c_EM) results. Falsely depressed Enzymatic Hemoglobin A1c results may affect consideration of intervention.

Product
Advia Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP), SMN 11319151
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov