RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink

Recalled January 21, 2015 · FDA recall Z-0987-2015 · Siemens Healthcare Diagnostics

Hazard
Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result could be erroneous and critical and appear to be verified upon repeat. · FDA Class II
Units
7
Sold at
Distributed in the state of WA.

Description

Enabling sending of preliminary/initial results on the ADVIA Chemistry XPT creates a risk that a critical result could be overwritten by the same result rather than showing the repeated result on CentraLink. A result could be erroneous and critical and appear to be verified upon repeat.

Product
ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink system software is a network solution provider and multi-system data manager for the instruments and lab automation systems (LAS) within the lab. The CentraLink software consolidates data from all connected instruments so that an operator can review and edit patient and quality control results from a single location.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Siemens Healthcare Diagnostics, Inc. recalls ADVIA Chemistry XPT interfaced to the CentraLink Data Management System V14x in specific configurations: The CentraLink | RecallWatchUSA