Siemens Healthcare Diagnostics, Inc. recalls Atellica CI Analyzer. Catalog Numbers: 10947347.
Recalled December 11, 2024 · FDA recall Z-0597-2025 · Siemens Healthcare Diagnostics
Hazard
Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior. · FDA Class II
Units
559 units
Sold at
Worldwide distribution - US Nationwide and the countries of Austria, Bangladesh, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Iraq, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Malaysia, Netherlands, Oman, Pakistan, Paraguay, Peru, Philippines, Portugal, Qatar, Republic Korea, Romania, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom Great Britain, Vietnam.
Description
Potential that the IMT Diluent volume remaining (% remaining) does not decrease as expected on the Atellica CI Analyzer, potentially leading to the IMT Diluent being empty while still displaying that volume is remaining. In this case Sodium (Na), Potassium (K) and Chloride (Cl) test results may be falsely elevated. Quality Control materials demonstrate the same behavior.
Product
Atellica CI Analyzer. Catalog Numbers: 10947347.