Siemens Healthcare Diagnostics, Inc. recalls Atellica IM 1300 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic test
Recalled November 25, 2020 · FDA recall Z-0473-2021 · Siemens Healthcare Diagnostics
Hazard
Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values · FDA Class II
Units
US 392 units; OUS 449 units
Sold at
Worldwide distribution - US Nationwide distribution and the countries of Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cura¿ao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, P.R. China, Pakistan, Peru, Poland, Portugal, Qatar Republic, Korea, Romania, Russian Fed., Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay, Vatikancity, Vietnam.
Description
Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values
Product
Atellica IM 1300 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →