RecallWatchUSA

Home / Recalls / Medical Devices

Siemens Healthcare Diagnostics, Inc. recalls Dimension¿ Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.

Recalled August 6, 2014 · FDA recall Z-2133-2014 · Siemens Healthcare Diagnostics

Hazard
There is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp. · FDA Class III
Units
15
Sold at
FL LA AR IA OH OH WI IL LA KS MT

Description

There is a misassembly issue with the photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.

Product
Dimension¿ Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Siemens Healthcare Diagnostics, Inc. recalls Dimension¿ Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids. | RecallWatchUSA