Siemens Healthcare Diagnostics, Inc. recalls Dimension Lipase Flex¿ Reagent Cartridge - ( Lipase-esterase, enzymatic, photometric, lipase)

Recalled May 16, 2018 · FDA recall Z-1796-2018 · Siemens Healthcare Diagnostics

Hazard
Calibration failure - confirmed a positive bias with LIPL lot FB8109 when compared to other Dimension LIPL lots. Lot FB8109 is not meeting Siemens internal specifications for lot-to-lot limits. · FDA Class II
Units
N/A
Sold at
Domestic and Foreign consignees

Description

Calibration failure - confirmed a positive bias with LIPL lot FB8109 when compared to other Dimension LIPL lots. Lot FB8109 is not meeting Siemens internal specifications for lot-to-lot limits.

Product
Dimension Lipase Flex¿ Reagent Cartridge - ( Lipase-esterase, enzymatic, photometric, lipase)
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov