Siemens Healthcare Diagnostics, Inc. recalls Dimension(R) Chemistry Wash (RD701), Siemens Material Number (SMN) 10445052 Product Usage: Used in the Dimension(R

Recalled January 1, 2014 · FDA recall Z-0561-2014 · Siemens Healthcare Diagnostics

Hazard
Siemens Healthcare Diagnostics received complaints of Chemistry Wash causing QC and patient sample result shifts on CTN1, LTN1, FT4, TSH, PBNP and LPBN when using Chemistry Wash lots RD23031, RD23111, RD23241, RD23311, RD23391, and RD23461. These lots shipped between 12/12/2012 and 3/22/2013. Internal investigation has confirmed shifts can be encountered in two scenarios: 1 - When an affected l · FDA Class II
Units
17,719
Sold at
US Nationwide and OUS.

Description

Siemens Healthcare Diagnostics received complaints of Chemistry Wash causing QC and patient sample result shifts on CTN1, LTN1, FT4, TSH, PBNP and LPBN when using Chemistry Wash lots RD23031, RD23111, RD23241, RD23311, RD23391, and RD23461. These lots shipped between 12/12/2012 and 3/22/2013. Internal investigation has confirmed shifts can be encountered in two scenarios: 1 - When an affected l

Product
Dimension(R) Chemistry Wash (RD701), Siemens Material Number (SMN) 10445052 Product Usage: Used in the Dimension(R) system with the heterogeneous module
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.