Siemens Healthcare Diagnostics, Inc. recalls Dimension Vista 500, Analyzer, chemistry Model no. 10284473

Recalled March 27, 2019 · FDA recall Z-1014-2019 · Siemens Healthcare Diagnostics

Hazard
When the Millipore Water Purification Module Progard Pretreatment Packs are used with Dimension Vista Triglyceride assay, patient and QC triglyceride results may be erroneously elevated as a result of contamination. · FDA Class II
Units
5373
Sold at
Nationwide domestic distribution. Worldwide foreign distribution.

Description

When the Millipore Water Purification Module Progard Pretreatment Packs are used with Dimension Vista Triglyceride assay, patient and QC triglyceride results may be erroneously elevated as a result of contamination.

Product
Dimension Vista 500, Analyzer, chemistry Model no. 10284473
Manufacturer(s)
Siemens Healthcare Diagnostics, Inc.